Impact of Drug-Coated Balloon Inflation Time on Outcome After Percutaneous Coronary Intervention : An Observational Study From SCAAR
(2026) In Journal of the American Heart Association 15(15).- Abstract
BACKGROUND: There are limited data guiding the optimal inflation time for coronary drug-coated balloons. The aim of this study was to assess the clinical efficacy of an inflation time of ≥30 seconds compared with <30 seconds.
METHODS: We conducted an observational nationwide analysis using the SCAAR (Swedish Coronary Angiography and Angioplasty Registry) to identify all patients undergoing percutaneous coronary intervention with drug-coated balloons and their inflation times. Outcomes were assessed at 1 year using Kaplan-Meier estimates and adjusted Cox regression.
RESULTS: The <30-second group comprised 412 patients, and the ≥30-second group 10 826 patients. Patients in the <30-second group presented more often... (More)
BACKGROUND: There are limited data guiding the optimal inflation time for coronary drug-coated balloons. The aim of this study was to assess the clinical efficacy of an inflation time of ≥30 seconds compared with <30 seconds.
METHODS: We conducted an observational nationwide analysis using the SCAAR (Swedish Coronary Angiography and Angioplasty Registry) to identify all patients undergoing percutaneous coronary intervention with drug-coated balloons and their inflation times. Outcomes were assessed at 1 year using Kaplan-Meier estimates and adjusted Cox regression.
RESULTS: The <30-second group comprised 412 patients, and the ≥30-second group 10 826 patients. Patients in the <30-second group presented more often with lesions in proximal segments. After a 1-year follow-up, ≥30 seconds was associated with lower rates of target segment revascularization compared with <30 seconds (3.2% versus 6.4%, adjusted hazard ratio [HR], 0.47 [95% CI, 0.32-0.69], P≤0.001) and target vessel myocardial infarction (1.8% versus 5.3%, adjusted HR, 0.32 [95% CI, 0.21-0.50], P≤0.001). No difference was observed for all-cause mortality.
CONCLUSIONS: In this analysis, including 11 238 patients undergoing PCI with drug-coated balloons, ≥30 second inflation time was associated with improved outcomes compared with a shorter inflation time.
(Less)
- author
- von Koch, Sacharias
LU
; Zhou, Mikael
LU
; Håkansson, Anton
LU
; Råmunddal, Truls
; Omerovic, Elmir
; Zwackman, Sammy
; Erlinge, David
LU
; Rissanen, Tuomas T
and Mohammad, Moman A
LU
- organization
- publishing date
- 2026-08-04
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- Humans, Female, Time Factors, Registries, Angioplasty, Balloon, Coronary/instrumentation, Treatment Outcome, Male, Aged, Coronary Artery Disease/therapy, Coronary Angiography, Coated Materials, Biocompatible, Middle Aged, Sweden/epidemiology, Cardiac Catheters, Cardiovascular Agents/administration & dosage, Risk Factors, Percutaneous Coronary Intervention/instrumentation, Myocardial Infarction
- in
- Journal of the American Heart Association
- volume
- 15
- issue
- 15
- article number
- e047354
- publisher
- Wiley-Blackwell
- external identifiers
-
- pmid:42535531
- scopus:105046669492
- ISSN
- 2047-9980
- DOI
- 10.1161/JAHA.125.047354
- language
- English
- LU publication?
- yes
- id
- 8695fc55-431d-43be-b389-ef4906264129
- date added to LUP
- 2026-08-18 14:46:10
- date last changed
- 2026-09-16 05:40:30
@article{8695fc55-431d-43be-b389-ef4906264129,
abstract = {{<p>BACKGROUND: There are limited data guiding the optimal inflation time for coronary drug-coated balloons. The aim of this study was to assess the clinical efficacy of an inflation time of ≥30 seconds compared with <30 seconds.</p><p>METHODS: We conducted an observational nationwide analysis using the SCAAR (Swedish Coronary Angiography and Angioplasty Registry) to identify all patients undergoing percutaneous coronary intervention with drug-coated balloons and their inflation times. Outcomes were assessed at 1 year using Kaplan-Meier estimates and adjusted Cox regression.</p><p>RESULTS: The <30-second group comprised 412 patients, and the ≥30-second group 10 826 patients. Patients in the <30-second group presented more often with lesions in proximal segments. After a 1-year follow-up, ≥30 seconds was associated with lower rates of target segment revascularization compared with <30 seconds (3.2% versus 6.4%, adjusted hazard ratio [HR], 0.47 [95% CI, 0.32-0.69], P≤0.001) and target vessel myocardial infarction (1.8% versus 5.3%, adjusted HR, 0.32 [95% CI, 0.21-0.50], P≤0.001). No difference was observed for all-cause mortality.</p><p>CONCLUSIONS: In this analysis, including 11 238 patients undergoing PCI with drug-coated balloons, ≥30 second inflation time was associated with improved outcomes compared with a shorter inflation time.</p>}},
author = {{von Koch, Sacharias and Zhou, Mikael and Håkansson, Anton and Råmunddal, Truls and Omerovic, Elmir and Zwackman, Sammy and Erlinge, David and Rissanen, Tuomas T and Mohammad, Moman A}},
issn = {{2047-9980}},
keywords = {{Humans; Female; Time Factors; Registries; Angioplasty, Balloon, Coronary/instrumentation; Treatment Outcome; Male; Aged; Coronary Artery Disease/therapy; Coronary Angiography; Coated Materials, Biocompatible; Middle Aged; Sweden/epidemiology; Cardiac Catheters; Cardiovascular Agents/administration & dosage; Risk Factors; Percutaneous Coronary Intervention/instrumentation; Myocardial Infarction}},
language = {{eng}},
month = {{08}},
number = {{15}},
publisher = {{Wiley-Blackwell}},
series = {{Journal of the American Heart Association}},
title = {{Impact of Drug-Coated Balloon Inflation Time on Outcome After Percutaneous Coronary Intervention : An Observational Study From SCAAR}},
url = {{http://dx.doi.org/10.1161/JAHA.125.047354}},
doi = {{10.1161/JAHA.125.047354}},
volume = {{15}},
year = {{2026}},
}