The Lund Nuclear Microprobe sub-micron set-up. Part II: Beam line, focusing system and scanning
(2005) In Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms 231(1-4). p.7-13- Abstract
- A new beam line for generating a sub-micron beam spot for high-resolution nuclear microprobe applications has been constructed at the Lund Nuclear Microprobe facility. The design and construction of the beam line, the quality test of the four high precision Oxford OM-52 Quadrupole magnets by the grid shadow method and the beam scanning system are presented in this paper. The beam line is designed for a two-stage focusing system using an independent doublet at each stage. The first stage focuses the beam in an intermediate chamber, which has a scanning system and a high resolution viewing system. The beam spot in the intermediate chamber is then used as the object for the second stage. Slit construction, vacuum system and beam control is... (More)
- A new beam line for generating a sub-micron beam spot for high-resolution nuclear microprobe applications has been constructed at the Lund Nuclear Microprobe facility. The design and construction of the beam line, the quality test of the four high precision Oxford OM-52 Quadrupole magnets by the grid shadow method and the beam scanning system are presented in this paper. The beam line is designed for a two-stage focusing system using an independent doublet at each stage. The first stage focuses the beam in an intermediate chamber, which has a scanning system and a high resolution viewing system. The beam spot in the intermediate chamber is then used as the object for the second stage. Slit construction, vacuum system and beam control is discussed. Data illustrating the quadrupole quality and scanning capability are shown. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/236021
- author
- Shariff, Asad LU ; Nilsson, Christer LU ; Auzelyte, Vaida LU ; Elfman, Mikael LU ; Kristiansson, Per LU ; Malmqvist, Klas LU ; Pallon, Jan LU and Wegdén, Marie LU
- organization
- publishing date
- 2005
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- sub-micron beam, quadrupole doublets, beam scanning, Nuclear Microprobe
- in
- Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms
- volume
- 231
- issue
- 1-4
- pages
- 7 - 13
- publisher
- Elsevier
- external identifiers
-
- wos:000229752400003
- scopus:28544433311
- ISSN
- 0168-583X
- DOI
- 10.1016/j.nimb.2005.01.027
- language
- English
- LU publication?
- yes
- additional info
- The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Nuclear Physics (Faculty of Technology) (011013007)
- id
- 2c61b55d-b875-427a-af18-0295f4141471 (old id 236021)
- date added to LUP
- 2016-04-01 16:49:28
- date last changed
- 2022-03-15 03:11:27
@article{2c61b55d-b875-427a-af18-0295f4141471, abstract = {{A new beam line for generating a sub-micron beam spot for high-resolution nuclear microprobe applications has been constructed at the Lund Nuclear Microprobe facility. The design and construction of the beam line, the quality test of the four high precision Oxford OM-52 Quadrupole magnets by the grid shadow method and the beam scanning system are presented in this paper. The beam line is designed for a two-stage focusing system using an independent doublet at each stage. The first stage focuses the beam in an intermediate chamber, which has a scanning system and a high resolution viewing system. The beam spot in the intermediate chamber is then used as the object for the second stage. Slit construction, vacuum system and beam control is discussed. Data illustrating the quadrupole quality and scanning capability are shown.}}, author = {{Shariff, Asad and Nilsson, Christer and Auzelyte, Vaida and Elfman, Mikael and Kristiansson, Per and Malmqvist, Klas and Pallon, Jan and Wegdén, Marie}}, issn = {{0168-583X}}, keywords = {{sub-micron beam; quadrupole doublets; beam scanning; Nuclear Microprobe}}, language = {{eng}}, number = {{1-4}}, pages = {{7--13}}, publisher = {{Elsevier}}, series = {{Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms}}, title = {{The Lund Nuclear Microprobe sub-micron set-up. Part II: Beam line, focusing system and scanning}}, url = {{http://dx.doi.org/10.1016/j.nimb.2005.01.027}}, doi = {{10.1016/j.nimb.2005.01.027}}, volume = {{231}}, year = {{2005}}, }