Particle flow rate from the airways as fingerprint diagnostics in mechanical ventilation in the intensive care unit : A randomised controlled study

Hallgren, Filip ; Stenlo, Martin LU ; Niroomand, Anna LU ; Broberg, Ellen LU ; Hyllén, Snejana LU ; Malmsjö, Malin LU and Lindstedt, Sandra LU (2021) In ERJ open research 7(3).
Abstract

Introduction Mechanical ventilation can be monitored by analysing particles in exhaled air as measured by particle flow rate (PFR). This could be a potential method of detecting ventilator-induced lung injury (VILI) before changes in conventional parameters can be detected. The aim of this study was to investigate PFR during different ventilation modes in patients without lung pathology. Method A prospective study was conducted on patients on mechanical ventilation in the cardiothoracic intensive care unit (ICU). A PExA 2.0 device was connected to the expiratory limb on the ventilator for continuous measurement of PFR in 30 patients randomised to either volume-controlled ventilation (VCV) or pressure-controlled ventilation (PCV) for 30... (More)

Introduction Mechanical ventilation can be monitored by analysing particles in exhaled air as measured by particle flow rate (PFR). This could be a potential method of detecting ventilator-induced lung injury (VILI) before changes in conventional parameters can be detected. The aim of this study was to investigate PFR during different ventilation modes in patients without lung pathology. Method A prospective study was conducted on patients on mechanical ventilation in the cardiothoracic intensive care unit (ICU). A PExA 2.0 device was connected to the expiratory limb on the ventilator for continuous measurement of PFR in 30 patients randomised to either volume-controlled ventilation (VCV) or pressure-controlled ventilation (PCV) for 30 min including a recruitment manoeuvre. PFR measurements were continued as the patients were transitioned to pressure-regulated volume control (PRVC) and then pressure support ventilation (PSV) until extubation. Results PRVC resulted in significantly lower PFR, while those on PSV had the highest PFR. The distribution of particles differed significantly between the different ventilation modes. Conclusions Measuring PFR is safe after cardiac surgery in the ICU and may constitute a novel method of continuously monitoring the small airways in real time. A low PFR during mechanical ventilation may correlate to a gentle ventilation strategy. PFR increases as the patient transitions from controlled mechanical ventilation to autonomous breathing, which most likely occurs as recruitment by the diaphragm opens up more distal airways. Different ventilation modes resulted in unique particle patterns and could be used as a fingerprint for the different ventilation modes.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
ERJ open research
volume
7
issue
3
article number
00961-2020
publisher
European Respiratory Society
external identifiers
  • scopus:85115690547
  • pmid:34322553
ISSN
2312-0541
DOI
10.1183/23120541.00961-2020
language
English
LU publication?
yes
additional info
Publisher Copyright: © The authors 2021.
id
892200d7-c7ee-4ef4-ab98-a2ba9389ff8f
date added to LUP
2021-11-01 13:13:45
date last changed
2024-06-15 19:33:42
@article{892200d7-c7ee-4ef4-ab98-a2ba9389ff8f,
  abstract     = {{<p>Introduction Mechanical ventilation can be monitored by analysing particles in exhaled air as measured by particle flow rate (PFR). This could be a potential method of detecting ventilator-induced lung injury (VILI) before changes in conventional parameters can be detected. The aim of this study was to investigate PFR during different ventilation modes in patients without lung pathology. Method A prospective study was conducted on patients on mechanical ventilation in the cardiothoracic intensive care unit (ICU). A PExA 2.0 device was connected to the expiratory limb on the ventilator for continuous measurement of PFR in 30 patients randomised to either volume-controlled ventilation (VCV) or pressure-controlled ventilation (PCV) for 30 min including a recruitment manoeuvre. PFR measurements were continued as the patients were transitioned to pressure-regulated volume control (PRVC) and then pressure support ventilation (PSV) until extubation. Results PRVC resulted in significantly lower PFR, while those on PSV had the highest PFR. The distribution of particles differed significantly between the different ventilation modes. Conclusions Measuring PFR is safe after cardiac surgery in the ICU and may constitute a novel method of continuously monitoring the small airways in real time. A low PFR during mechanical ventilation may correlate to a gentle ventilation strategy. PFR increases as the patient transitions from controlled mechanical ventilation to autonomous breathing, which most likely occurs as recruitment by the diaphragm opens up more distal airways. Different ventilation modes resulted in unique particle patterns and could be used as a fingerprint for the different ventilation modes.</p>}},
  author       = {{Hallgren, Filip and Stenlo, Martin and Niroomand, Anna and Broberg, Ellen and Hyllén, Snejana and Malmsjö, Malin and Lindstedt, Sandra}},
  issn         = {{2312-0541}},
  language     = {{eng}},
  month        = {{07}},
  number       = {{3}},
  publisher    = {{European Respiratory Society}},
  series       = {{ERJ open research}},
  title        = {{Particle flow rate from the airways as fingerprint diagnostics in mechanical ventilation in the intensive care unit : A randomised controlled study}},
  url          = {{http://dx.doi.org/10.1183/23120541.00961-2020}},
  doi          = {{10.1183/23120541.00961-2020}},
  volume       = {{7}},
  year         = {{2021}},
}